Nuclear Medicine PETCT Anti-Radiation Dispensing fume hood
Protective fume hood (glove box)
Adopting a fully isolated protection design with an independent air supply and exhaust system, it is used to meet the needs of radioactive drug rinsing, packaging, and labeling.
The protective fume hood consists of three parts, with the control area located at the upper part. It is equipped with an independent exhaust unit unit and a 15 degree inward tilt control panel for easy daily use and observation by the operator; The work area is located in the middle, with lead glass windows and two operator hand holes installed on the front for entering the work area; The auxiliary area is located at the lower part and is equipped with an activity meter ionization chamber shielding well and a molybdenum technetium generator shielding transport device, fully meeting the functional requirements for the use of radioactive drugs in leaching, packaging, and labeling processes.
The fume hood provides high standards of radiation protection, ensuring that operators comply with ICRP radiation safety environmental standards during the process of rinsing, packaging, and labeling radioactive drugs.
Work areas that meet cGMP biological cleanliness standards
Integrated stainless steel work area with vertical laminar negative pressure ventilation system
Air purification system with optional self circulation or external exhaust
Cleanliness level: Class A (ISO 100 level) (optional)
Air supply method: Vertical laminar flow (optional)
Laminar flow velocity: 0.35-0.56 m/sec (optional)
Required air volume: 800 m ³/h
Sealing method: isolated clean circular arc angle stainless steel inner cavity
Supply air filtration: Fully efficient HEPA 99.995% @ 0.5 µ m (H13)
Exhaust filtration: adsorption of radioactive gas filter
Lead shielding that meets ICRP radiation environmental safety standards